FDA Adverse Event Other Summary report: N

GMK FEMORAL CUTTING GUIDE 4INL SIZE 4

MDR report key: 3723689 · Received December 19, 2013

Report

Report Number
3005180920-2013-00178
Event Type
Other
Date Received
December 19, 2013
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FROM THE DOCUMENT REVIEW OF THE LOT INVOLVED (1112890 - (B)(4) ITEMS PRODUCED) NO ANOMALIES WERE FOUND. THE CUTTING GUIDE AND THE (B)(4) PINS RETURNED BACK TO MEDACTA AND WERE INSPECTED BY QC DEPT WITH THE FOLLOWING RESULTS: THE CUTTING GUIDE DOES NOT PRESENT EVIDENCES OF WEAR AND ALL THE FIXATION HOLES ARE FUNCTIONING, W/O METALLIC FRAGMENTS. THE PINS HAVE A FEW SUPERFICIAL SCRATCHES CAUSED BY NORMAL USE BUT THERE IS NO DETACHMENT OF MATERIAL AND RESULTING TO BE FUNCTIONING. THE ONLY POSSIBLE EXPLANATION IS THAT THE FRAGMENTS CAME FROM THE THREADED DRILL USED TO PERFORM THE PRELIMINARY HOLES PRIOR TO FIX THE PINS.

Description of Event or Problem · 1

REF IMP # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666272 GMK FEMORAL CUTTING GUIDE 4INL SIZE 4 REUSABLE SURGICAL INSTRUMENT FOR KNEE LZO MEDACTA INTERNATIONAL SA

Patients

Seq Age Sex Outcome Treatment
1