FDA Adverse Event Malfunction Summary report: N

FISHER & PAYKEL 850 RESPIRATORY HUMIDIFICATION SYSTEM

MDR report key: 371747 · Received December 20, 2001

Report

Report Number
MW4003173
Event Type
Malfunction
Date Received
December 20, 2001
Date of Event
November 2, 2001
Report Date
December 18, 2001
Manufacturer
FISHER & PAYKEL HEALTH CARE INC.
Product Code
BTT
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

UPON NORMAL EXTENDED SELF TEST (EST) OF THE PURITAN BENNETT 7200 VENTILATOR AN ERROR CODE WAS REALIZED REPEATIVELY. BIO-MED WAS SUMMONED AND MADE THE REPAIR OF THE VENT WHICH WAS A Q-3 SENSOR AND BOARD. AGAIN A DIFFERENT VENTILATOR HAS THE SAME PROBLEM. BIO-MED WAS CALLED AGAIN. THEY SAID WATER WAS GETTING INTO THE MACHINE. THE ONLY DIFFERENCE IS THAT HOSP HAS A NEW HUMIDIFICATION SYSTEM ON THE VENTILATORS. THEY CALLED PURITAN-BENNETT AND THE TECH SUPPORT PERSON STATED THAT THE 7200 WAS NOT DESIGNED TO ACCOMMODATE A HEATED EXPIRATORY LIMB ON THERE CIRCUITS. THE EXPIRED AIR IS HEATED AND THE FIRST THING COLD IT ENCOUNTERS IS THE Q-3 SENSOR, WHERE THE HEATED VAPOR RAINS OUT (DEW POINT) AND GETS THE INSIDE OF THE MACHINE WET. THE CIRCUIT ALSO INVOLVED AS PART OF THAT SYSTEM IS THE AIRLIFE CIRCUIT TM, CAT# RT110, MFG BY FISHER & PAYKEL AND DISTRIBUTED BY ALLEGIANCE. NO PTS WERE AFFECTED OR KNOWN TO BE AFFECTED. NO ERROR CODES OR MALFUNCTIONS WERE NOTED DURING PT USE. THE MACHINES WERE TAGGED AND SECURED UNTIL A SOLUTION WAS DEVELOPED WITH THE ASSISTANCE AND SUGGESTIONS FROM BOTH PB TECHNICAL SUPPORT AND FISHER & PAYKEL CLINICAL SPECIALIST. NO PROBLEMS HAVE BEEN NOTED SINCE A NON-HEATED EXPIRATORY LIMB HAS BEEN IMPLIMENTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58545 FISHER & PAYKEL 850 RESPIRATORY HUMIDIFICATION SYSTEM HUMIDIFIER BTT FISHER & PAYKEL HEALTH CARE INC. 850 NA

Patients

Seq Age Sex Outcome Treatment
1 NA PURITAN BENNETT 7200 VENTILATOR, 2001| AIRLIFE CIRCUIT TM CAT# RT110, 2001