FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 3717267
·
Received April 1, 2014
Report
- Report Number
- 3004753838-2014-18044
- Event Type
- Malfunction
- Date Received
- April 1, 2014
- Date of Event
- January 15, 2013
- Report Date
- January 15, 2013
- Manufacturer
- DEXCOM INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION OF THE RETURNED RECEIVER AND REVIEW OF THE DATA LOG CONFIRMED THE PERMANENT OUT OF RANGE SIGNAL. DATA WAS RECEIVED (B)(6) 2013. A TRANSMITTER (SN (B)(4), MFG DATE: 12/04/2012) WAS ALSO RETURNED FOR EVALUATION AND THERE WAS NO ISSUE DETECTED.
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION OF THE RETURNED DEVICE AND REVIEW OF THE DATA LOG DID NOT CONFIRM THE REPORTED PERMANENT LOSS OF "Y".
Description of Event or Problem · 1
CONSUMER REPORTED PERMANENT OUT OF RANGE SIGNAL. THE CONSUMER DID NOT REPORT ANY MEDICAL INTERVENTION OR INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 195880 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS, PRODUCT CODE:MDS | MDS | DEXCOM INC. | MT20649 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |