FDA Adverse Event
Summary report: N
INTELLISPACE PORTAL
MDR report key: 3716700
·
Received March 4, 2014
Report
- Report Number
- 1525965-2014-00040
- Date Received
- March 4, 2014
- Report Date
- February 5, 2014
- Manufacturer
- PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
- Product Code
- LLZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). WE WILL FILE A F/U MDR AT THE COMPLETION OF THE INVESTIGATION. INTERNAL CROSS REFERENCE: COMPLAINT (B)(4)
Description of Event or Problem · 1
PHILIPS REC'D FEEDBACK FROM A CUSTOMER, REPORTING THAT THERE ARE DIFFERENCES IN THE QUANTITATIVE PERFUSION SPECT (QPS) IN AUTOQUANT RESULTS BETWEEN THE INTELLISPACE PORTAL (ISP) AND EXTENDED BRILLIANCE WORKSPACE (EBW). IT WAS FOUND THAT THE INTELLISPACE PORTAL RECONSTRUCTION SOMETIMES PRODUCES INACCURATE RESULTS WHEN IT USES THE ASTONISH ALGORITHM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 129171 | INTELLISPACE PORTAL | LLZ | PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. | 881001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening |