OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-07754
- Event Type
- Malfunction
- Date Received
- April 1, 2014
- Report Date
- March 24, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
FOLLOW-UP # 1 - 7/9/2014. DEVICE EVALUATION:THE LAY USER/PATIENT¿S PRODUCTS HAVE BEEN RETURNED ON 6/2/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 7/7/2014 WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE RESULTS. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿161, 170, 172, 181, AND 177 MG/DL¿ WITH A LIFESCAN METER AND ¿149, 118, 122, 138, 136, AND 108 MG/DL¿ ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193266 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |