FDA Adverse Event Malfunction Summary report: N

SYNTHES (USA) ELECTRIC PEN DRIVE SYSTEM

MDR report key: 3714758 · Received March 31, 2014

Report

Report Number
3009450871-2014-10050
Event Type
Malfunction
Date Received
March 31, 2014
Date of Event
March 3, 2014
Report Date
March 3, 2014
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PK043310
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: CORRECTION: THE MANUFACTURER LOCATION WAS DOCUMENTED AS WALDENBURG IN THE INITIAL REPORT. THE LOCATION HAS BEEN UPDATED TO (B)(4). CONTACT OFFICE NAME/ADDRESS HAVE BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECTED MANUFACTURING FACILITY. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION.  RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED.  THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE NORMAL WEAR FROM USE AND SERVICING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSPECTION, THE MOTOR DEVICE DID ¿NOT WORK WITH ANY CONSOLE". THERE WAS NO PATIENT INVOLVEMENT. THE REPORTER STATED THAT THE EVENT DID NOT OCCUR DURING SURGERY AS THE EQUIPMENT WAS USED FOR DEMONSTRATION PURPOSES ONLY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
189035 SYNTHES (USA) ELECTRIC PEN DRIVE SYSTEM INSTR, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR/ACCESS & ATTACH HWE SYNTHES OBERDORF 1029

Patients

Seq Age Sex Outcome Treatment
1 CONSOLE DEVICE