FDA Adverse Event
Death
Summary report: N
PRECISION
MDR report key: 371463
·
Received January 16, 2002
Report
- Report Number
- 371463
- Event Type
- Death
- Date Received
- January 16, 2002
- Date of Event
- January 11, 2002
- Report Date
- January 16, 2002
- Manufacturer
- PRECISION MEDICAL INC
- Product Code
- CAX
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
PT WAS ELECTIVELY INTUBATED AFTER THE CONCLUSION OF A CARDIAC CATH PROCEDURE AND RECEIVED NITROUS OXIDE INSTEAD OF OXYGEN. THE OXYGEN FLOW METER WAS USED WITH AN ADAPTER/CONNECTOR. ONE OF TWO PRONGS WAS BROKEN OFF CONNECTOR, ALLOWING INSERTION INTO NITROUS OXIDE PORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION | OXYGEN FLOW METER AND CONNECTOR | CAX | PRECISION MEDICAL INC | IMFA4001-FLOW METER | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Death |