FDA Adverse Event Death Summary report: N

PRECISION

MDR report key: 371461 · Received January 16, 2002

Report

Report Number
371461
Event Type
Death
Date Received
January 16, 2002
Date of Event
January 15, 2002
Report Date
January 16, 2002
Manufacturer
PRECISION MEDICAL INC
Product Code
CAX
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

OXYGEN FLOW METER USED WITH ADAPTER/CONNECTOR. ONE OF TWO PRONGS BROKEN OFF CONNECTOR, ALLOWING INSERTION INTO NITROUS OXIDE PORT. PT WAS ON VENTILATOR DURING CODE SITUATION OCCURRING DURING CARDIAC CATH, AND RECEIVED NITROUS OXIDE INSTEAD OF OXYGEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION OXYGEN FLOW METER AND CONNECTOR CAX PRECISION MEDICAL INC 1MFA4001-FLOWMETER *

Patients

Seq Age Sex Outcome Treatment
1 68 YR Death