FDA Adverse Event
Other
Summary report: N
SMOKE EVACUATOR
MDR report key: 37141
·
Received July 3, 1996
Report
- Report Number
- 37141
- Event Type
- Other
- Date Received
- July 3, 1996
- Date of Event
- March 5, 1996
- Report Date
- March 11, 1996
- Manufacturer
- IC MEDICAL, INC.
- Product Code
- FYD
- Report Source
- User Facility report
- Reporter Location
- MI, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
SMOKE EVACUATOR WAS BEING USED FOR A BREAST REDUCTION FOR APPROX ONE HR. RPTR SAW SMOKE AND RED SPARKS COMING OUT OF THE FRONT OF THE MACHINE. A HOLE QUICKLY BURNED THROUGH THE MACHINE. MACHINE WAS TURNED OFF AND REMOVED FROM THE OR. BIO-MED REPORT: THIS UNIT IS A DEMO. THE FRONT PANEL MELTED NEAR THE FILTER CONNECTOR. IT HAS A RIBBON CABLE SUPPLYING POWER TO A FRONT PANEL WARNING "SINGLE USE FILTERS." "FLOW" SET AT MAX; "TIME" SET AT 1/4 CW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SMOKE EVACUATOR | SMOKE EVACUATION SYSTEM | FYD | IC MEDICAL, INC. | 350 D | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR | Other |