Description of Event or Problem · 1

SMOKE EVACUATOR WAS BEING USED FOR A BREAST REDUCTION FOR APPROX ONE HR. RPTR SAW SMOKE AND RED SPARKS COMING OUT OF THE FRONT OF THE MACHINE. A HOLE QUICKLY BURNED THROUGH THE MACHINE. MACHINE WAS TURNED OFF AND REMOVED FROM THE OR. BIO-MED REPORT: THIS UNIT IS A DEMO. THE FRONT PANEL MELTED NEAR THE FILTER CONNECTOR. IT HAS A RIBBON CABLE SUPPLYING POWER TO A FRONT PANEL WARNING "SINGLE USE FILTERS." "FLOW" SET AT MAX; "TIME" SET AT 1/4 CW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SMOKE EVACUATOR SMOKE EVACUATION SYSTEM FYD IC MEDICAL, INC. 350 D *

Patients

Seq Age Sex Outcome Treatment
1 31 YR Other