FDA Adverse Event
Other
Summary report: N
HILL-ROM
MDR report key: 370553
·
Received January 7, 2002
Report
- Report Number
- 1836145-2002-00001
- Event Type
- Other
- Date Received
- January 7, 2002
- Date of Event
- November 2, 2001
- Manufacturer
- HILL-ROM ARCHITECTURAL PRODUCTS HILL-ROM CO. INC.
- Product Code
- FNJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
INFO RECEIVED FROM INDICATED THAT A NURSE WAS INJURED ON THE CHEST AND BOTH ARMS AS THE NURSE ATTEMPTED TO CATCH THE MONITOR (65 LBS). FURTHER INVESTIGATION SHOWS THAT THE WRONG SCREWS WAS USED TO INSTALL THE MONITOR CHANNEL TO THE MONITOR CLAMPS ASSY. A 3/4 INCH SCREWS WAS USED INSTEAD OF THE 1 INCH CALLED OUT ON THE DRAWINGS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HILL-ROM | MONITOR MOUNTING CHANNEL ASSEMBLY | FNJ | HILL-ROM ARCHITECTURAL PRODUCTS HILL-ROM CO. INC. | P100705 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |