FDA Adverse Event Other Summary report: N

HILL-ROM

MDR report key: 370553 · Received January 7, 2002

Report

Report Number
1836145-2002-00001
Event Type
Other
Date Received
January 7, 2002
Date of Event
November 2, 2001
Manufacturer
HILL-ROM ARCHITECTURAL PRODUCTS HILL-ROM CO. INC.
Product Code
FNJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

INFO RECEIVED FROM INDICATED THAT A NURSE WAS INJURED ON THE CHEST AND BOTH ARMS AS THE NURSE ATTEMPTED TO CATCH THE MONITOR (65 LBS). FURTHER INVESTIGATION SHOWS THAT THE WRONG SCREWS WAS USED TO INSTALL THE MONITOR CHANNEL TO THE MONITOR CLAMPS ASSY. A 3/4 INCH SCREWS WAS USED INSTEAD OF THE 1 INCH CALLED OUT ON THE DRAWINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HILL-ROM MONITOR MOUNTING CHANNEL ASSEMBLY FNJ HILL-ROM ARCHITECTURAL PRODUCTS HILL-ROM CO. INC. P100705 *