B LOCUS HIGH RES SSP UNITRAY KIT
Report
- Report Number
- 2244574-2014-00077
- Event Type
- Malfunction
- Date Received
- January 15, 2014
- Report Date
- May 31, 2012
- Manufacturer
- LIFE TECHNOLOGIES CORP
- Product Code
- MZI
- PMA / PMN Number
- BK020068
- Removal / Correction Number
- 006-2012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
INTERNAL INVESTIGATION HAS CONFIRMED THAT PRIMER MIX B-086 FOUND ON LANE 9 IN THE B LOCUS HIGH RES SSP UNITRAY KIT (CATALOG #4730010 LOT #009 910766) WILL GIVE A FALSE NEGATIVE FOR THE B*47:03 ALLELE IN THE AFFECTED PRODUCT ACCORDING TO THE LABELING. THE LABELING STATED THE PRIMERS IN PRIMER MIX B-086 WOULD AMPLIFY THIS SEQUENCE IN THE B*47:03 ALLELE. ROOT CAUSE HAS BEEN DETERMINED TO BE PRODUCT DESIGN (INCORRECT LABELING).
IT WAS REPORTED IN CUSTOMER COMPLAINT # (B)(4) THAT A SURVEY SAMPLE TESTED WHEN USING B LOCUS HIGH RES SSP UNITRAY KIT (CATALOG #4730010, LOT #009 984375) RESULTED IN B*47:02 WHEN THE RESULT SHOULD HAVE BEEN B*47:03 ((B)(4)). CATALOG #4730010 B LOCUS HIGH RES SSP UNITRAY KIT, LOT #009 910766 HAS BEEN IDENTIFIED AS PRODUCT THAT WOULD BE AFFECTED BY THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41136 | B LOCUS HIGH RES SSP UNITRAY KIT | MZI, TEST,QUALITATIVE,FOR HLA,NON-DIAGNOSTIC | MZI | LIFE TECHNOLOGIES CORP | 009 910766 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |