DRB1*15 SSP UNITRAY KIT WITH TAQ POLYMERASE
Report
- Report Number
- 2244574-2014-00016
- Event Type
- Malfunction
- Date Received
- January 9, 2014
- Date of Event
- October 28, 2013
- Report Date
- October 28, 2013
- Manufacturer
- LIFE TECHNOLOGIES CORPORATION
- Product Code
- MZI
- PMA / PMN Number
- BK020068
- Removal / Correction Number
- 013-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
AN INVESTIGATION WAS COMPLETED TO VERIFY HOW THE PRIMER REACTIVITIES ASSIGNED WITHIN THE INTERNAL LEGACY SCORE DATABASE WERE INTERPRETED WITHIN THE CURRENT HOS (HUMAN OLIGOTYPING SOFTWARE) DATABASE. THE INVESTIGATION SHOWED THERE WERE NO OTHER INCORRECT PRIMER REACTIVITIES, HOWEVER, THE INVESTIGATION IDENTIFIED THAT SOME OF THE PRIMER REACTIVITY ASSIGNMENTS WERE INCONSISTENT BETWEEN PRIMER MIXES. THE INCONSISTENCY IS DUE TO NEW ALLELIC INFO BEING AVAILABLE NOW, WHICH WAS NOT AVAILABLE AT THE TIME WHEN THE INFO WAS INITIALLY TRANSFERRED FROM THE SCORE DATABASE TO HOS DATABASE IN 2008. THROUGH THIS INVESTIGATION, ONE OF THE PRIMER REACTIVITIES THAT WAS IDENTIFIED AS NEEDING HARMONIZATION HAD REACTION PATTERNS ON THE DATE INTERPRETATION WORKSHEETS AND UNIMATCH SOFTWARE THAT MAY BE INTERPRETED AS A POTENTIAL MIS-TYPE ASSIGNMENT FOR DRB1X15:34 ALLELE AS DRB1X16:01:02 AND DRB1X15:54/66 ALLELE AS DRB1X15:25. AS A PRODUCT MAINTENANCE FUNCTION, THE PRIMER REACTIVITIES WERE UPDATED IN THE HOS DATABASE AND ON THE DATA INTERPRETATION WORKSHEETS AND UNITMATCH SOFTWARE. A CUSTOMER NOTIFICATION WAS ISSUED TO INFORM CUSTOMERS OF THE CHANGE TO THE REACTIVITY PATTERS SHOWN ON THE LABELING WORKSHEETS AND UNIMATCH SOFTWARE.
A DISCREPANCY IN TYPING RESULTS USING DRB1X15 SSP UNITRAY KIT (CATALOG # 450102D, LOT 011 1274159) WAS REPORTED IN INTERNAL COMPLAINT #(B)(4). IT WAS DISCOVERED THAT PRIMER MIX R15-05C, WHICH IS INCLUDED IN LANE 5 OF DRB1X15 SSP UNITRAY KIT (CATALOG #450102D, LOT 011 1274159, HAS BEEN SHOWN TO PRODUCE A NEGATIVE REACTION IN THE PRESENCE OF THE DRB1X15:34/54/66 ALLELES. AS A RESULT OF THE NEGATIVE REACTION A SAMPLE CONTAINING DRB1X15:34 ALLELE COULD MISTYPE AS A DRB1X16:01:02 AND A SAMPLE CONTAINING DRB1X15:54/66 ALLELE COULD MISTYPE AS DRB1X15:25. THE INVESTIGATION OF INTERNAL COMPLAINT #(B)(4) IDENTIFIED DRB1X15 SSP UNITRAY KIT WITH TAQ POLYMERASE, CATALOG # 450112D, LOT #011 1049955 1273533 AS A PRODUCT AFFECTED BY THE SAME ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 17228 | DRB1*15 SSP UNITRAY KIT WITH TAQ POLYMERASE | MZI TEST, QUALITATIVE FOR HLA, NON-DIAGN | MZI | LIFE TECHNOLOGIES CORPORATION | 0111049955 1273533 |