FDA Adverse Event
Malfunction
Summary report: N
MEDICAL DATA ELECTRONICS (MDE)
MDR report key: 370021
·
Received December 17, 2001
Report
- Report Number
- 370021
- Event Type
- Malfunction
- Date Received
- December 17, 2001
- Date of Event
- December 1, 2001
- Report Date
- December 17, 2001
- Manufacturer
- MDE
- Product Code
- DRT
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PT BEING MONITORED AFTER CONSCIOUS SEDATION. MONITOR BEGAN READING SVT OR V-FIB. PT FOUND TO BE IN NO DISTRESS. PLEASE TAKE NOTE THAT STRIP INDICATED HR OF 82 BEATS PER MINUTES. PT NOT INJURED. STAFF TREATED BASED ON CLINICAL FINDINGS. STRIP REPRINTOUT OF EVENT WITH NO VF/SVT. NURSE QUESTIONS CELL PHONE INTERFERENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56873 | MEDICAL DATA ELECTRONICS (MDE) | ESCORT VISION CENTRAL STATION | DRT | MDE | VISION-20500-16 | 1.3.0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR | Other | SOFTWARE UPGRADE PROVIDED 12/06/2001 TO CLIMINATE| PROBLEM. |