FDA Adverse Event Other Summary report: N

ARJO POWERED AMBULIFT

MDR report key: 36984 · Received August 28, 1996

Report

Report Number
1922538-1996-00082
Event Type
Other
Date Received
August 28, 1996
Date of Event
August 19, 1996
Report Date
August 28, 1996
Manufacturer
ARJO MANUFACTURING CO.
Product Code
FSA
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE FOLLOWING IS A CLOSING OUT REPORT BY THE FOREIGN REPORTER'S INVESTIGATOR. INVESTIGATION: THE LIFT WAS EXAMINED BUT NO FAULTS COULD BE FOUND. THIS LIFT HAS BEEN USED IN THE SAME ROOM WITH A HILO BATH MANY TIMES BEFORE. OBSERVATIONS: THE INVESTIGATOR HAS OFFERED A POSSIBLE EXPLANATION OF EVENTS - THE JIB WAS LOWERED TOO FAR, THE CHAIR TOUCHED THE BOTTOM OF THE BATH AND MOVED AWAY FROM THE JIB LOCATION. WHEN THE LIFT WAS RAISED THE CHAIR WAS NOT LOCATED PROPERLY ON THE JIB AND THE MOVEMENT OF THE CHAIR RE-ALIGNMENT ITS SELF. FACILITY AGREED THIS WAS POSSIBLE. CONCLUSIONS: USER ERROR. HOIST LOWERED TO FAR. CORRECTIVE ACTION: HOSIT HAS BEEN PUT BACK INTO SERVICE, AND STAFF NOT UNDERSTANDS THE PROBLEMS IT THE HOIST IS LOWERED TOO FAR.

Description of Event or Problem · 1

THE CHAIR TIPPED AS STAFF WAS LOWERING A FEMALE PT INTO THE BATH. SHE WAS MOBILE ENOUGH TO GET OFF THE CHAIR. NO INJURY TO PT OR STAFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARJO POWERED AMBULIFT POWERED BATHING HOIST FSA ARJO MANUFACTURING CO. ABXXXX NA

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other