FDA Adverse Event Malfunction Summary report: N

ADM BALL IMPACTOR TIP

MDR report key: 3693164 · Received March 21, 2014

Report

Report Number
0002249697-2014-00870
Event Type
Malfunction
Date Received
March 21, 2014
Date of Event
February 26, 2014
Report Date
February 26, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING DAMAGE INVOLVING AN ADM IMPACTOR WAS REPORTED. THE EVENT WAS CONFIRMED FRACTURE OCCURRED IN MULTIPLE OVERLOADS, FRACTURING FROM THE INSIDE DIAMETER OUT. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE SURFACES EXAMINED. ALL DEVICES IN THE REPORTED LOT WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THERE HAVE BEEN NO OTHER EVENTS FOR THE LOT REFERENCED. THIS EVENT WAS PREVIOUSLY INVESTIGATED AND IS CURRENTLY UNDER A CAPA INVESTIGATION. IT WAS DETERMINED THAT THERE IS A DESIGN ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING MDM SURGERY, WHEN A SURGEON WAS IMPACTED MDM LINER WITH THE REPORTED DEVICE, IT BROKE.

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING MDM SURGERY, WHEN A SURGEON WAS IMPACTED MDM LINER WITH THE REPORTED DEVICE, IT BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
169704 ADM BALL IMPACTOR TIP INSTRUMENT LXH STRYKER ORTHOPAEDICS-MAHWAH G3654009

Patients

Seq Age Sex Outcome Treatment
1 Other