ADM BALL IMPACTOR TIP
Report
- Report Number
- 0002249697-2014-00870
- Event Type
- Malfunction
- Date Received
- March 21, 2014
- Date of Event
- February 26, 2014
- Report Date
- February 26, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
AN EVENT REGARDING DAMAGE INVOLVING AN ADM IMPACTOR WAS REPORTED. THE EVENT WAS CONFIRMED FRACTURE OCCURRED IN MULTIPLE OVERLOADS, FRACTURING FROM THE INSIDE DIAMETER OUT. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE SURFACES EXAMINED. ALL DEVICES IN THE REPORTED LOT WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THERE HAVE BEEN NO OTHER EVENTS FOR THE LOT REFERENCED. THIS EVENT WAS PREVIOUSLY INVESTIGATED AND IS CURRENTLY UNDER A CAPA INVESTIGATION. IT WAS DETERMINED THAT THERE IS A DESIGN ISSUE.
IT WAS REPORTED THAT, DURING MDM SURGERY, WHEN A SURGEON WAS IMPACTED MDM LINER WITH THE REPORTED DEVICE, IT BROKE.
IT WAS REPORTED THAT, DURING MDM SURGERY, WHEN A SURGEON WAS IMPACTED MDM LINER WITH THE REPORTED DEVICE, IT BROKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169704 | ADM BALL IMPACTOR TIP | INSTRUMENT | LXH | STRYKER ORTHOPAEDICS-MAHWAH | G3654009 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |