FDA Adverse Event
Malfunction
Summary report: N
AQUACISER
MDR report key: 36920
·
Received September 6, 1996
Report
- Report Number
- MW1009874
- Event Type
- Malfunction
- Date Received
- September 6, 1996
- Date of Event
- August 14, 1996
- Report Date
- August 20, 1996
- Manufacturer
- FERNO-WASHINGTON, INC.
- Product Code
- ILJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- TX, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
2/96 - FACILITY NOTIFIED VERBALLY BY MFR OF POSSIBILITY OF CORROSION OF HOSE CONNECTIONS AND RESULTING LEAK OF HYDRAULIC FLUID. 3/23/96 CO REP AT FACILITY TO REPAIR UNIT AND TAKE CARE OF HOSE CONNECTIONS. REP DECIDED THAT HOSE CONNECTIONS WERE NOT A PROBLEM AND DID NOT REPLACE THEM. 8/14/96 LEAK OF ALL THE HYDRAULIC FLUID (3 GALLONS) INTO POOL. MSDS STATES THAT FLUID CAN CAUSE SKIN OR EYE PROBLEMS. NO KNOWN PT PROBLEMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AQUACISER | HYDRO-MASSAGE BATH | ILJ | FERNO-WASHINGTON, INC. | 976 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |