FDA Adverse Event Malfunction Summary report: N

MAR COR COMPLETE WATER TREATMENT SYSTEM FOR KIDNEY DIALYSIS

MDR report key: 3691048 · Received March 20, 2014

Report

Report Number
3019131-2014-00010
Event Type
Malfunction
Date Received
March 20, 2014
Date of Event
July 31, 2013
Report Date
March 20, 2014
Manufacturer
MAR COR PURIFICATION
Product Code
FIP
PMA / PMN Number
K945559
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CARBON TANK HEADS SPLIT/ CRACKED CAUSING AND LEAKED WATER ALL OVER THE RO ROOM AND PATIENT DIALYSIS WAS DELAYED. MAR COR PURIFICATION SERVICE REPRESENTATIVE WENT TO FACILITY AND REPAIRED AND REPLACED BROKEN PARTS OF SYSTEM. PLEASE REFERENCE ATTACHED PICTURES. TO THIS DATE, THERE WERE NO REPORTS OF PATIENT INJURY OR OTHER ADVERSE PATIENT IMPACT; HOWEVER, DIALYSIS TREATMENT WAS DELAYED. THIS COMPLAINT WILL CONTINUE TO BE MONITORED WITHIN MCP COMPLAINT SYSTEM. IF ADDITIONAL INFORMATION IS RECEIVED, THIS CASE WILL BE REVIEWED UPON RECEIPT.

Description of Event or Problem · 1

CARBON TANK HEADS THREADED CONNECTION CRACKED AND LEAKED WATER ALL OVER THE RO ROOM OF THE DIALYSIS CLINIC CAUSING DELAY IN PATIENT TREATMENT.