OT ULTRALINK METER
Report
- Report Number
- 2939301-2014-06414
- Event Type
- Injury
- Date Received
- March 19, 2014
- Date of Event
- March 9, 2014
- Report Date
- March 12, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K073231
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL TESTS WERE PERFORMED ON THE DESICCANT TO CHECK FOR POSSIBLE MOISTURE ISSUE. THE DESICCANT WITHIN THE SUBJECT VIAL WAS FOUND TO BE SATURATED BY MOISTURE AND FAILED TGA TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
FOLLOW-UP # 1 ¿ (05/26/2014), THE PATIENT¿S METER AND TEST STRIPS HAVE BEEN RETURNED ON (B)(4) 2014 AND (B)(4) 2014, RESPECTIVELY, AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE TEST STRIPS WERE FOUND TO HAVE RESULTS GREATER THAN +50% OF THE CONTROL SOLUTION WHEN TESTED WITH CONTROL SOLUTION. ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL TO CHECK THE STRUCTURAL INTEGRITY. THE STRUCTURAL INTEGRITY OF THE TEST STRIP VIAL WAS FOUND TO BE INTACT DURING A GROSS LEAK TEST. A DHR (DEVICE HISTORY RECORD) WAS COMPLETED ON (B)(4) 2014 FOR THIS PRODUCT AND NO DEVIATIONS, NON-CONFORMANCES, OR REWORKS WERE OBSERVED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2014, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HIS ONETOUCH ULTRALINK METER READ INACCURATELY HIGH AND ERRATIC. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON THE MORNING OF (B)(6) 2014. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿200, 215 AND 220 MG/DL¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS DO NOT EXCEED THE EXPECTED VALUE OF <=20% AND/ OR <=20 MG/DL. THE PATIENT MANAGES HIS DIABETES WITH INSULIN PUMP THERAPY. IN RESPONSE TO THE ALLEGED RESULTS, THE PATIENT REPORTEDLY INCREASED HIS USUAL DOSE OF NOVOLOG INSULIN. ABOUT 20 MINUTES AFTER, THE PATIENT CLAIMED HE FELT SYMPTOMS OF NERVOUS AND SHAKY. NO TREATMENT WAS SPECIFIED. THE PATIENT DENIED TESTING WITH ANOTHER DEVICE. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT. THE PATIENT WAS TESTING FROM UNAPPROVED SAMPLE SITES AT THE TIME OF THE CONCERN. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO PERFORM QUALITY CONTROL TESTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT CLAIMED HE OBTAINED INACCURATE READINGS ON THE SUBJECT METER, ADMINISTERED TREATMENT BASED ON THE ALLEGED RESULTS, AND REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163005 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3464837 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Life Threatening |