FDA Adverse Event Malfunction Summary report: N

12.6CM TAPERED ATTACH, BM

MDR report key: 3686580 · Received March 18, 2014

Report

Report Number
1045834-2014-11584
Event Type
Malfunction
Date Received
March 18, 2014
Date of Event
February 27, 2014
Report Date
March 3, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
GFF
PMA / PMN Number
PK974025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE UNIT WAS TESTED WHICH FOUND THE TEMPERATURE WAS ABOVE SPECIFICATION. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR OF COMPONENT PARTS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(6). THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SPINE DECOMPRESSION/FUSION SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS ¿GETTING VERY HOT WHEN ON AND ACTIVATED¿. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
160341 12.6CM TAPERED ATTACH, BM BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT GFF DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR