LUMIWAVE INFRARED THERAPY DEVICE
Report
- Report Number
- 1055581-2006-00041
- Event Type
- Malfunction
- Date Received
- June 26, 2006
- Date of Event
- June 2, 2006
- Report Date
- June 26, 2006
- Manufacturer
- BIOCARE SYSTEMS, INC.
- Product Code
- ILY
- Removal / Correction Number
- CAR# 101
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THIS STANDARD IS USED BY UL WHEN TESTING FOR UL 60950-1 APPROVAL FOR WHICH THIS PRODUCT CLAIMS COMPLIANCE. FOR THIS ET OF TESTS, EACH UNIT WAS TESTED IN ACCORDANCE WITH SECTION 43 DIRECT PLUG-IN BLADE SECURENESS TEST OF THE UL 1310 SPECIFICATION. THIS TEST REQUIRED A 20LB PULL TEST FOR 2 MINUTES ON EACH INDIVIDUAL BLADE AND THEN ON BOTH BLADES TOGETHER WITH THE UNIT SUPPORTED HORIZONTALLY. THE 20LB WEIGHT USED WAS MEASURED AT 20.35LBS ON A CALIBRATED SCALE. ALL UNITS PASSED UL 1310'S DIRECT PLUG-IN BLADE SECURENESS TEST. IN ADDITION TO TESTING THE PRODUCT TO THE UL STANDARD 1310 SECTION 43, A PULL TEST WAS PERFORMED VIA THE CORD WITH THE AC ADAPTER PLUGGED INTO A WALL OUTLET. THIS TEST WAS DONE TO REPLICATE THE INFORMATION PROVIDED FROM THE ORIGINAL COMPLAINT AND ALTHOUGH IT IS CONSIDERED A MISUSE APPLICATION, IT DOES REFLECT A POTENTIAL REAL WORLD SCENARIO. FOR THIS TEST, EACH UNIT WAS PLUGGED INTO AN ORDINARY WALL OUTLET AND PULLED OUT BY THE CORD. THE FORCE USED TO PULL THE UNITS OUT WAS NOT EXCESSIVE (I.E. THE UNIT WAS NOT RIPPED FROM THE WALL VIOLENTLY). THE CORD WAS PULLED WITH A MODERATE, FIRM JERK THAT WAS ENOUGH TO REMOVE THE AC ADAPTER FROM THE WALL OUTLET. EVERY UNIT FAILED ON THE FIRST PULL. THE UNITS THAT WERE SUBJECTED TO A STRAIGHT OR VIRTUALLY STRAIGHT PULL FAILED WITH ONE BLADE BEING COMPLETELY REMOVED FROM THE UNIT AND THE SECOND BLADE BEING EITHER VERY LOOSE MECHANICALLY OR HANGING/DANGLING OUT OF THE AC ADAPTER HOUSING. THE UNITS SUBJECTED TO A 45 DEGREES ANGULAR PULL FAILED WITH BOTH BLADES COMPLETELY REMOVED FROM THE AC ADAPTER. (B)(4).
USER WAS UTILIZING THE DEVICE ON ELBOW WHILE SITTING IN A SWIVEL CHAIR AT HIS DESK. HW SWIVELED IN THE CHAIR, PULLING THE AC ADAPTER FROM OUTLET, WHICH LEFT THE PRONGS BEHIND IN THE WALL SOCKET. NO INJURY OR ELECTRICAL SHOCK OCCURRED AS A RESULT OF THIS EVENT. PRODUCT PROBLEM AND PRODUCT USE ERROR: DURING USE OF THE PRODUCT, USER PULLED THE POWER SUPPLY FROM THE WALL OUTLET BY THE POWER CORD AND NOT THE POWER SUPPLY HOUSING. THE RESULTING FAILURE WAS THAT THE PRONGS OF THE POWER SUPPLY BECAME SEPARATED FORM THE HOUSING AND REMAINED IN THE WALL-OUTLET. IT IS NOTED THAT THE COMPONENT FAILURE IS FOR AN OFF-THE SHELF ITEM USED IN MANY CONSUMER PRODUCTS. THERE HAVE BEEN A TOTAL OF THREE REPORTED INCIDENTS OF THIS FAILURE MODE AT THIS TIME. THE PRODUCT POWER SUPPLY WAS VISUALLY INSPECTED. IT WAS CONFIRMED THAT THE PRONGS FROM THE POWER SUPPLY WERE DETACHED FROM THE SWAGING LOCATION. OBVIOUS DEFORMATION OF PLASTIC SURROUNDING THE ENTRY POINT OF THE BLADES TO THE HOUSING INDICATES THE UNIT WAS PULLED FROM THE WALL BY THE CORD. THE POWER SUPPLY ITSELF IS A COMMERCIALLY AVAILABLE PRODUCT; AC ADAPTER MODEL KSAFE1200200W1US PART # DPS120200U-P5P-TK MANUFACTURED BY CUI.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LUMIWAVE INFRARED THERAPY DEVICE | INFRARED LAMP | ILY | BIOCARE SYSTEMS, INC. | LUMIWAVE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |