FDA Adverse Event Injury Summary report: N

RESTORE ULTRA

MDR report key: 3673483 · Received March 11, 2014

Report

Report Number
3004209178-2014-04270
Event Type
Injury
Date Received
March 11, 2014
Report Date
February 20, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 37711, SERIAL# (B)(4), IMPLANTED: (B)(6) 2006, PRODUCT TYPE: IMPLANTABLE NEUROSTIMULATOR; PRODUCT ID 377860, LOT# V011437, IMPLANTED: (B)(6) 2006, PRODUCT TYPE: LEAD; PRODUCT ID 37742, SERIAL# (B)(4), IMPLANTED: (B)(6) 2006, PRODUCT TYPE: PROGRAMMER, PATIENT; PRODUCT ID 37752, SERIAL# (B)(4), IMPLANTED: (B)(6) 2006, PRODUCT TYPE: RECHARGER; PRODUCT ID 377860, LOT# V011437, IMPLANTED: (B)(6) 2006, PRODUCT TYPE: LEAD; PRODUCT ID 37743, SERIAL# (B)(4), PRODUCT TYPE: PROGRAMMER, PATIENT; PRODUCT ID 3708120, SERIAL# (B)(4), IMPLANTED: (B)(6) 2009, PRODUCT TYPE: EXTENSION; PRODUCT ID 3777-45, SERIAL# (B)(4), IMPLANTED: (B)(6) 2009, PRODUCT TYPE: LEAD; PRODUCT ID 3708120, SERIAL# (B)(4), IMPLANTED: (B)(6) 2009, PRODUCT TYPE: EXTENSION; PRODUCT ID 3778-45, SERIAL# (B)(4), IMPLANTED: (B)(6) 2009, PRODUCT TYPE: LEAD; PRODUCT ID 37752, SERIAL# (B)(4), PRODUCT TYPE: RECHARGER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THERE WAS AN IMPEDANCE READING OF >3600 OHMS. IT WAS NOTED THAT EVERY SINGLE ELECTRODE PAIR WAS SHOWING GREATER THAN 3600 OHMS, BUT PATIENT WAS STILL FEELING STIMULATION. IT WAS NOTED THE PATIENTS DEVICE CURRENTLY FELT "GOOD IN THE LEGS". IT WAS REPORTED THAT ONLY ONE IMPEDANCE TEST WAS PERFORMED AT .24V FOR THE DEVICE. IT WAS REPORTED THE SYSTEM HAD HIGH IMPEDANCES. IT WAS REPORTED THE PATIENT WAS PROBABLY GOING TO GET SYSTEM REPLACED WITH ANOTHER DEVICE SINCE HE HAD MS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
145260 RESTORE ULTRA STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37712

Patients

Seq Age Sex Outcome Treatment
1 00059 YR Required Intervention