COBAS E-ELECSYS TESTOSTERONE GEN.2
Report
- Report Number
- 1823260-2014-01660
- Event Type
- Malfunction
- Date Received
- March 10, 2014
- Date of Event
- February 11, 2014
- Report Date
- March 25, 2014
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- CDZ
- PMA / PMN Number
- K093421
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
THE REAGENT LOT IN USE AT THE TIME OF EVENT WAS 171423. THE EXPIRATION DATE WAS REQUESTED, BUT WAS NOT PROVIDED. A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. ADDITIONAL INFORMATION FOR FURTHER INVESTIGATION WAS REQUESTED BUT WAS NOT PROVIDED. REVIEW OF THE PROVIDED CALIBRATION AND QC DATA DID NOT INDICATE A GENERAL REAGENT ISSUE. INSTRUMENT PERFORMANCE CHECKS DID NOT PRODUCE ANY CONCERNS WITH PERFORMANCE OF THE ANALYZER.
THE CUSTOMER RECEIVED A QUESTIONABLE TESTOSTERONE RESULT FOR ONE PATIENT SAMPLE FROM THE COBAS E601 ANALYZER SERIAL NUMBER (B)(4). TWO SPECIMENS WERE DRAWN FROM THE PATIENT. THE TESTOSTERONE RESULT FROM THE FIRST TUBE WAS 663.6 NG/DL. THE REPEAT RESULT WAS 237.2 NG/DL. THE RESULT FROM THE SECOND TUBE WAS 232.2 NG/DL. THE RESULT OF 663.6 NG/DL WAS RELEASED OUTSIDE THE LABORATORY. THE REPEAT RESULT OF 237.2 NG/DL WAS BELIEVED TO BE CORRECT. THERE WAS NO HARM TO THE PATIENT. THE FIELD SERVICE REPRESENTATIVE COULD NOT FIND A CAUSE. HE NOTED REPLACEMENT OF THE REAGENT DID NOT RESOLVE THE PROBLEM. HE RAN PERFORMANCE TESTING ON THE ANALYZER AND PERFORMED EXTENSIVE TROUBLE SHOOTING OVER SEVERAL DAYS WHICH SHOWED NO PROBLEMS. SEVERAL PRECISION STUDIES GAVE POOR RESULTS AND INDICATED REAGENT INSTABILITY THAT COULD PRODUCE QUESTIONABLE RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142939 | COBAS E-ELECSYS TESTOSTERONE GEN.2 | RADIOIMMUNOASSAY, TESTOSTERONES AND DIHYDROTESTOSTERONE | CDZ | ROCHE DIAGNOSTICS | NA | 171423 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 057 YR |