12.6CM TAPERED ATTACH, BM
Report
- Report Number
- 1045834-2014-11592
- Event Type
- Malfunction
- Date Received
- March 7, 2014
- Date of Event
- February 27, 2014
- Report Date
- February 27, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- GFF
- PMA / PMN Number
- PK974025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
THIS DEVICE WAS RETURNED FOR SERVICE HOWEVER DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE NORMAL WEAR ON THE BEARINGS OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE (B)(4) EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE HAD "HIGH TEMPERATURE". THIS EVENT WAS NOT RELATED TO SURGERY. THERE WERE NO INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 140490 | 12.6CM TAPERED ATTACH, BM | BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT | GFF | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |