DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE
Report
- Report Number
- 2244574-2014-00166
- Event Type
- Malfunction
- Date Received
- January 23, 2014
- Report Date
- June 13, 2012
- Manufacturer
- LIFE TECHNOLOGIES CORP
- Product Code
- MZI
- PMA / PMN Number
- BK070038
- Removal / Correction Number
- 004-2012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
INTERNAL INVESTIGATION CONFIRMS CUSTOMER INCIDENT OF LANE 95 BASE PAIR SELECTION NOT LIMITING RESULTS TO EITHER DR4 OR DR5 BASE ON SELECTION OF 85B[ PR 210 BP RESULT. THIS IS A KNOWN DEFECT OF THE UNIMATCH SOFTWARE WHEN USING A UCH FILE WERE THE "^" SYMBOLS ARE PRESENT IN THE DRB4/DRB5 MIX. THE PRESENCE OF THE "^" SYMBOL BREAKS THE SIZE FILTERING FUNCTION OF UNIMATCH WHICH RESULTS IN BOTH DRB4 AND DRB5 DISPLAYED IN THE RESULTS REGARDLESS OF THE PRODUCT SIZE ENTERED BY THE USER. DCO1035227 RELEASED 4/06/2012 CONTAINED NEW UCH FILES UPDATING THE REACTIVITY OF DRB1*04:33 IN LANE 44 OF THE DRB1 KIT. THE UCH FILES RELEASED IN DC01035227 WERE DONE SO WITHOUT THE REQUIRED MANUAL EDITS TO REMOVE THE "^ " SYMBOLS IN LANE 95. DISPOSITION IS TO USE AS IS. A CUSTOMER NOTIFICATION WAS SENT ON 05/31/2012 TO AFFECTED CUSTOMERS PROVIDING INSTRUCTIONS ON HOW TO OBTAIN CORRECTED DOCUMENTS FROM THE WEBSITE. ROOT CAUSE IS DETERMINED TO BE AN ERROR IN THE MANUAL PROCESS OF CREATING THIS FUNCTIONALITY WHEN GENERATING THE KIT DATABASE IN HOSII.
IT WAS REPORTED THAT DURING TESTING USING THE ALLSET GOLD SSP HLA-DRB1 HIGH RES KIT (CATALOG NUMBER: 54040D LOT #: 017 904113) THAT THE DRB1 LANE 95 BASE PAIR FILTER DID NOT FUNCTION PROPERLY. RESULTS WERE NOT LIMITED TO EITHER DR4 OR DR5. THIS WOULD RESULT IN A MISTYPE. ((B)(4)). 4515110 DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE, LOT # 017 904113 1025404 HAS BEEN IDENTIFIED AS ALSO HAVING THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54636 | DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE | MZI TEST, QUALITATIVE FOR HLA, NON-DIAGNOSTIC | MZI | LIFE TECHNOLOGIES CORP | 017 904113 1025404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |