FDA Adverse Event Malfunction Summary report: N

DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE

MDR report key: 3664004 · Received January 23, 2014

Report

Report Number
2244574-2014-00171
Event Type
Malfunction
Date Received
January 23, 2014
Report Date
June 13, 2012
Manufacturer
LIFE TECHNOLOGIES CORP
Product Code
MZI
PMA / PMN Number
BK070038
Removal / Correction Number
004-2012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL INVESTIGATION CONFIRMS CUSTOMER INCIDENT OF LANE 95 BASE PAIR SELECTION NOT LIMITING RESULTS TO EITHER DR4 OR DR5 BASE ON SELECTION OF 85B[ PR 210 BP RESULT. THIS IS A KNOWN DEFECT OF THE UNIMATCH SOFTWARE WHEN USING A UCH FILE WERE THE "^" SYMBOLS ARE PRESENT IN THE DRB4/DRB5 MIX. THE PRESENCE OF THE "^" SYMBOL BREAKS THE SIZE FILTERING FUNCTION OF UNIMATCH WHICH RESULTS IN BOTH DRB4 AND DRB5 DISPLAYED IN THE RESULTS REGARDLESS OF THE PRODUCT SIZE ENTERED BY THE USER. (B)(4) RELEASED (B)(4) 2012 CONTAINED NEW UCH FILES UPDATING THE REACTIVITY OF DRB1*04:33 IN LANE 44 OF THE DRB1 KIT. THE UCH FILES RELEASED IN (B)(4) WERE DONE SO WITHOUT THE REQUIRED MANUAL EDITS TO REMOVE THE "^ " SYMBOLS IN LANE 95. DISPOSITION IS TO USE AS IS. A CUSTOMER NOTIFICATION WAS SENT (B)(4) 2012 TO AFFECTED CUSTOMERS PROVIDING INSTRUCTIONS ON HOW TO OBTAIN CORRECTED DOCUMENTS FROM THE WEBSITE. ROOT CAUSE IS DETERMINED TO BE AN ERROR IN THE MANUAL PROCESS OF CREATING THIS FUNCTIONALITY WHEN GENERATING THE KIT DATABASE IN HOSII.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING USING THE ALLSET GOLD SSP HLA-DRB1 HIGH RES KIT (CATALOG NUMBER: 54040D LOT #: 017 904113) THAT THE DRB1 LANE 95 BASE PAIR FILTER DID NOT FUNCTION PROPERLY. RESULTS WERE NOT LIMITED TO EITHER DR4 OR DR5. THIS WOULD RESULT IN A MISTYPE. ((B)(4)) 4515110 DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE, LOT # 017 904113 1083735 HAS BEEN IDENTIFIED AS ALSO HAVING THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54693 DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE MZI TEST, QUALITATIVE FOR HLA, NON-DIAGNOSTIC MZI LIFE TECHNOLOGIES CORP 017 904113 1096905

Patients

Seq Age Sex Outcome Treatment
1