FDA Adverse Event
Death
Summary report: N
THORACENTSIS/PARACENTESIS KIT
MDR report key: 365704
·
Received December 14, 2001
Report
- Report Number
- 1423507-2001-00286
- Event Type
- Death
- Date Received
- December 14, 2001
- Report Date
- December 14, 2001
- Manufacturer
- ALLEGIANCE HEALTHCARE CORP.
- Product Code
- GCB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
1. STYLET TOO LONG - ONCE IN CHEST CAVITY, CAN BE TOO CLOSE TO ORGANS; 2. STYLET BLUNT SO DOCTOR USED FORCE TO PUSH IT THROUGH THE CHEST WALL - ORGANS AT RISK. FURTHER COMMUNICATION REVEALED THAT PACK WAS USED ON A PT FOR A PLEURAL TAP PT WENT INTO CARDIAC ARREST 20 MINS LATER, AND THEY WERE UNABLE TO RESUSCITATE. AN AUTOPSY SHOWED PERFORATION OF THE LEFT VENTRICLE WITH CARDIAC TAMPONADE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56591 | THORACENTSIS/PARACENTESIS KIT | THORACENTSIS TRAY | GCB | ALLEGIANCE HEALTHCARE CORP. | TPK1001 | LIB037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Death |