FDA Adverse Event Death Summary report: N

THORACENTSIS/PARACENTESIS KIT

MDR report key: 365704 · Received December 14, 2001

Report

Report Number
1423507-2001-00286
Event Type
Death
Date Received
December 14, 2001
Report Date
December 14, 2001
Manufacturer
ALLEGIANCE HEALTHCARE CORP.
Product Code
GCB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

1. STYLET TOO LONG - ONCE IN CHEST CAVITY, CAN BE TOO CLOSE TO ORGANS; 2. STYLET BLUNT SO DOCTOR USED FORCE TO PUSH IT THROUGH THE CHEST WALL - ORGANS AT RISK. FURTHER COMMUNICATION REVEALED THAT PACK WAS USED ON A PT FOR A PLEURAL TAP PT WENT INTO CARDIAC ARREST 20 MINS LATER, AND THEY WERE UNABLE TO RESUSCITATE. AN AUTOPSY SHOWED PERFORATION OF THE LEFT VENTRICLE WITH CARDIAC TAMPONADE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56591 THORACENTSIS/PARACENTESIS KIT THORACENTSIS TRAY GCB ALLEGIANCE HEALTHCARE CORP. TPK1001 LIB037

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Death