FDA Adverse Event Other Summary report: N

MULTISTIX 10SG STRIPS

MDR report key: 3649068 · Received February 14, 2014

Report

Report Number
1217157-2014-00020
Event Type
Other
Date Received
February 14, 2014
Date of Event
January 15, 2014
Report Date
January 17, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JIO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER IS BEING REQUESTED TO SEND REAGENT BACK FOR FURTHER INVESTIGATION. CUSTOMER INDICATED THAT THEY ARE NOT CURRENTLY USING SIEMENS TEST STRIPS. THE CAUSE FOR THE DISCORDANT RESULTS IS UNKNOWN.

Additional Manufacturer Narrative · 1

THE INITIAL MDR 1217157-2014-00020 WAS FILED WITH FDA ON FEBRUARY 14, 2014. ADDITIONAL INFORMATION: BASED ON THE ADDITIONAL INFORMATION, SIEMENS TECHNICAL OPERATIONS TESTED CUSTOMER'S RETURNED REAGENT AND COULD NOT CONFIRM COMPLAINT. AS PER THE MODERN URINE CHEMISTRY MANUAL, "A NEGATIVE NITRITE SHOULD NEVER BE INTERPRETED AS INDICATING AN ABSENCE OF BACTERIURIA. THERE ARE SEVERAL REASONS FOR THIS. FIRST-MORNING URINE, OR URINE THAT HAS REMAINED IN THE BLADDER FOR FOUR OR MORE HOURS, IS MOR LIKELY TO YIELD A POSITIVE NITRITE TEST RESULT IN THE PRESENCE OF SIGNIFICANT BACTERIURIA THAN A RANDOM URINE SAMPLE THAT HAVE BEEN IN THE BLADDER ONLY A SHORT TIME. IN THE LATTER TYPE OF SPECIMEN, THERE MAY HAVE BEEN INSUFFICIENT TIME FOR THE CONVERSION OF NITRATE TO NITRITE BY THE INFECTING BACTERIA. SOME STRAINS OF URINARY PATHOGENS DO NOT PRODUCE THE ENZYME NECESSARY TO REDUCE NITRATE TO NITRITE, BUT THES ORGANISMS ARE THE LEAST COMMON PATHOGENS. A NEGATIVE TEST RESULT MAY ALSO OCCUR IF DIETARY NITRATES ARE ABSENT." CUSTOMER HAS BEEN PROVIDED WITH A REPLACEMENT BOTTLE OF REAGENT.

Description of Event or Problem · 1

CUSTOMER REPORTED NEGATIVE NITRITE AND LEUKOCYTES RESULTS ON THE INSTRUMENT WHEREAS CULTURE RESULTS WERE POSITIVE FOR BOTH ANALYTES. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97552 MULTISTIX 10SG STRIPS MULTISTIX 10SG JIO SIEMENS HEALTHCARE DIAGNOSTICS INC.

Patients

Seq Age Sex Outcome Treatment
1 13 YR