FDA Adverse Event Malfunction Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 3645810 · Received February 25, 2014

Report

Report Number
1030489-2014-00585
Event Type
Malfunction
Date Received
February 25, 2014
Date of Event
February 2, 2014
Report Date
February 2, 2014
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
NKB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFO: THE BURR THAT CAME OFF WAS NOT RETURNED. MICROSCOPIC EXAMINATION DOES NOT REVEAL ANY AREA THAT LOOKED CHIPPED. WITHOUT THE BURR NO FURTHER ANALYSIS CAN BE DONE.

Additional Manufacturer Narrative · 1

(B)(4): THIS PART IS NOT APPROVED FOR USE IN THE UNITED STATES; HOWEVER A LIKE DEVICE CATALOG # 5440030, 510K # K102555 WAS CLEARED IN THE UNITED STATES. NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE. IT WAS REPORTED THAT A METAL BURR WAS NOTICED COMING FROM THE SET SCREW. THE FRAGMENT WAS RETRIEVED AND THE SET SCREW WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116725 CD HORIZON SPINAL SYSTEM ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE NKB MDT SOFAMOR DANEK PUERTO RICO MFG NA H13J4138