FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT

MDR report key: 3642876 · Received February 24, 2014

Report

Report Number
1826988-2014-00051
Event Type
Malfunction
Date Received
February 24, 2014
Date of Event
February 6, 2014
Report Date
February 6, 2014
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K110894
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONTOUR TEST STRIPS ARE ALSO EXPECTED TO BE RETURNED FOR EVALUATION: LOT# 2KC3F08, EXP DATE 10/31/2014, MANUFACTURE DATE: 10/01/2012, 510(K): K110587. THE MODEL NUMBERS WERE NOT PROVIDED. IN SOME COUNTRIES OUTSIDE THE US, CUSTOMER INFORMATION IS NOT PROVIDED DUE TO PRIVACY LAWS.

Description of Event or Problem · 1

A (B)(6) CUSTOMER RAN BLOOD GLUCOSE TESTS ON TWO CONTOUR NEXT LINK METERS AND RECEIVED READINGS OF 287 AND 250MG/DL. THE CUSTOMER RETESTED ON A CONTOUR LINK METER AND THE READING WAS 120MG/DL. THE DIFFERENCES BETWEEN THE READINGS FALL IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCES CLINICALLY SIGNIFICANT NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. NEW METERS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114058 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE LLC 3HFED02

Patients

Seq Age Sex Outcome Treatment
1