FDA Adverse Event Malfunction Summary report: N

FLYBOOK CONSOLE WITH PRINTER

MDR report key: 3640019 · Received January 3, 2014

Report

Report Number
3006795936-2014-00001
Event Type
Malfunction
Date Received
January 3, 2014
Date of Event
December 27, 2013
Report Date
December 27, 2013
Manufacturer
VASONOVA INC.
Product Code
OBJ
PMA / PMN Number
K103260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT IS REPORTED THIS EVENT OCCURRED ON A FEMALE PATIENT WITH THE INSERTION SITE BEING THE RIGHT ARM. WHILE ACCESSING THE PATIENT USING THE FLYBOOK CONSOLE, THE CLINICIAN HAD CONCERN OVER WHICH VESSEL WAS THE ARTERY AND WHICH WAS THE VEIN BECAUSE THEY WERE EXTREMELY CLOSE TOGETHER. SHE STUCK WHAT SHE BELIEVED WAS THE VEIN, HAVING THE CONFIDENCE TO MOVE FORWARD WITH THE PROCEDURE BECAUSE SHE WAS USING VPS. THE CASE, IN HER ESTIMATION, WAS FAIRLY STANDARD AFTER THE INITIAL STICK AND VPS SHOWED A GREEN ARROW UNTIL A BLUE BULLS-EYE WAS OBTAINED. HOWEVER, DURING THE POST-INSERTION CHEST X-RAY PROCESS, THE RADIOLOGISTS COULDN'T CONFIRM WHETHER THE PICC WAS IN THE SVC OR IN AN ARTERY. STILL FEELING UNEASY, BLOOD GASES WERE DRAWN VIA RESPIRATORY THERAPY. THE READINGS WERE CONSISTENT WITH AN ARTERY STICK. THE PICC WAS REMOVED AND REPLACED. THE SECOND PICC WAS CONFIRMED AS ACCURATELY PLACED BY THE RADIOLOGY GROUP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4413 FLYBOOK CONSOLE WITH PRINTER CATHETER, ULTRASOUND, INTRAVASCULAR OBJ VASONOVA INC.

Patients

Seq Age Sex Outcome Treatment
1