FDA Adverse Event Malfunction Summary report: N

STATSPIN EXPRESS 2 CENTRIFUGE

MDR report key: 3639659 · Received December 31, 2013

Report

Report Number
2023446-2013-00047
Event Type
Malfunction
Date Received
December 31, 2013
Date of Event
September 20, 2013
Report Date
December 31, 2013
Manufacturer
IRIS INTERNATIONAL
Product Code
JQC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

STATSPIN CENTRIFUGE REPORTING LATCH UNLOCKS AND RELEASES WHEN THE STOP BUTTON IS PRESSED, WHICH CAN STILL RELEASE INNER CONTENTS. NO INJURIES REPORTED. THE INSTRUMENT WAS REPAIRED AND WAS VERIFIED TO BE OPERATIONAL VIA INSTRUMENT PERFORMANCE TESTS WITH 120V AND 1 HOUR BURN WITH NO ERRORS.

Description of Event or Problem · 1

CUSTOMER REPORTS THAT THE LATCH OPENS WHEN THE STOP BUTTON IS SELECTED. NO INJURIES WERE REPORTED.