FDA Adverse Event
Malfunction
Summary report: N
STATSPIN EXPRESS 2 CENTRIFUGE
MDR report key: 3639659
·
Received December 31, 2013
Report
- Report Number
- 2023446-2013-00047
- Event Type
- Malfunction
- Date Received
- December 31, 2013
- Date of Event
- September 20, 2013
- Report Date
- December 31, 2013
- Manufacturer
- IRIS INTERNATIONAL
- Product Code
- JQC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
STATSPIN CENTRIFUGE REPORTING LATCH UNLOCKS AND RELEASES WHEN THE STOP BUTTON IS PRESSED, WHICH CAN STILL RELEASE INNER CONTENTS. NO INJURIES REPORTED. THE INSTRUMENT WAS REPAIRED AND WAS VERIFIED TO BE OPERATIONAL VIA INSTRUMENT PERFORMANCE TESTS WITH 120V AND 1 HOUR BURN WITH NO ERRORS.
Description of Event or Problem · 1
CUSTOMER REPORTS THAT THE LATCH OPENS WHEN THE STOP BUTTON IS SELECTED. NO INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682830 | STATSPIN EXPRESS 2 CENTRIFUGE | STATSPIN EXPRESS 2 | JQC | IRIS INTERNATIONAL | EXPRESS 2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |