VENTED AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2014-00138
- Event Type
- Malfunction
- Date Received
- February 18, 2014
- Date of Event
- December 24, 2013
- Report Date
- January 30, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED AND CONNECTED TO A WATERBAG TO TEST FOR LEAKS. OUR INVESTIGATION IS ALSO BASED ON OUR PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS. RESULTS: SMALL DROPS OF WATER BEGAN TO BUILD UP AT THE CONNECTION BETWEEN THE WATER FEEDSET TUBE AND SPIKE. VISUAL INSPECTION REVEALED THAT SUFFICIENT AMOUNT OF GLUE WAS PRESENT AROUND THE SPIKE TUBING CONNECTION; HOWEVER, THE GLUE HAD ONLY PARTLY BONDED. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 121208. CONCLUSION: WE WERE UNABLE TO DETERMINE DEFINITIVELY THE ROOT CAUSE OF THE REPORTED FAULT; HOWEVER, IT IS POSSIBLE THAT THE LEAKS FROM THE FEEDSET TUBE WAS DUE TO THE FAILURE OF THE GLUE BOND THAT JOINS THE SPIKE TO THE WATER FEEDSET TUBE. OUR PREVIOUS INVESTIGATION ON SIMILAR COMPLAINTS HAVE ALSO SHOWN THAT SUCH PROBLEM HAS BEEN CAUSED BY THE USER REMOVING THE SPIKE BY GRASPING THE TUBING INSTEAD OF THE SPIKE. WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE AT THE CHAMBER DOME IN ANY OF OUR TESTING. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. ADDITIONALLY ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THIS SUGGESTS THAT THE PROBLEM OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 AUTOFEED HUMIDIFICATION CHAMBERS STATE THE FOLLOWING: - "SET APPROPRIATE VENTILATOR ALARM." - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT WATER LEAKED AT THE CONNECTION BETWEEN THE WATER FEEDSET TUBE AND WATER BAG SPIKE OF AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER DURING USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 102789 | VENTED AUTOFEED HUMIDIFICATION CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 1212080105 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT |