FDA Adverse Event Malfunction Summary report: N

BABY CONTROL COSYCOT INFANT WARMER

MDR report key: 3630827 · Received February 17, 2014

Report

Report Number
9611451-2014-00129
Event Type
Malfunction
Date Received
February 17, 2014
Date of Event
January 20, 2014
Report Date
January 20, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
FMT
PMA / PMN Number
K971695
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT DEVICE IS NOT EXPECTED TO BE RETURNED TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. THEREFORE, OUR ANALYSIS IS BASED ON THE PHOTOGRAPH PROVIDED BY THE MEDICAL CENTRE, INVESTIGATION OF SIMILAR COMPLAINTS AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: THE PHOTOGRAPH PROVIDED BY THE MEDICAL CENTRE REVEALED A LONGITUDINAL CRACK ON THE PLASTIC LEG NEAR THE BASE FRAME OF THE REPORTED IW934 COSYCOT INFANT WARMER. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 070730. CONCLUSION: THE CRACK DAMAGE TO THE LEG OF THE IW934 COSYCOT INFANT WARMER WAS MOST LIKELY DUE TO OVERLOADING OF THE DEVICE AND EXCEEDING THE MAXIMUM WEIGHT CAPACITY OF 115 KGS. TO INCREASE AWARENESS AND TO REDUCE THE INCIDENCE OF CRACKED BASES, A WARNING IS PROVIDED IN THE FORM OF A "115 KG MAX WEIGHT" LABEL APPLIED TO THE COLUMN OF THE INFANT WARMER. THE MAXIMUM WEIGHT LIMIT IS SPECIFIED IN BOTH THE TECHNICAL MANUAL AND THE OPERATING MANUAL OF THE INFANT WARMER.

Description of Event or Problem · 1

A MEDICAL CENTRE IN (B)(6) REPORTED THAT THE PLASTIC LEGS OF AN IW934JEU BABY CONTROL COSYCOT INFANT WARMER WERE BROKEN. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
100418 BABY CONTROL COSYCOT INFANT WARMER FMT FMT FISHER & PAYKEL HEALTHCARE LIMITED IW934JEU 070730

Patients

Seq Age Sex Outcome Treatment
1