INFANT EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00125
- Event Type
- Malfunction
- Date Received
- February 13, 2014
- Date of Event
- January 9, 2014
- Report Date
- January 14, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT267 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K103767. METHOD: TWO RT267 INFANT EVAQUA2 BREATHING CIRCUITS INCLUDING THE SWIVEL WYE-PIECE WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(6) FOR EVALUATION. THEY WERE VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: VISUAL INSPECTION REVEALED A CRACK ON THE SWIVEL WYE PIECE ON BOTH OF THE CIRCUITS. FURTHERMORE, ONE OF THE SWIVEL WYE PIECES WAS NOT PROPERLY SEATED IN THE ELBOW. PRESSURE TEST REVEALED THAT THIS SWIVEL WYE PIECE EXHIBITED LEAK AND WAS OUT OF SPECIFICATION. THE OTHER SWIVEL WYE WAS WITHIN SPECIFICATION. UPON CORRECTLY FITTING THE SWIVEL WYE PIECE INTO THE ELBOW THE SWIVEL WYE WAS PRESSURE TESTED AGAIN AND WAS WITHIN SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 121005. CONCLUSION: THE RETURNED SWIVEL WYE PIECES WERE MOST LIKELY SUBJECTED TO PHYSICAL FORCE SUFFICIENT TO CRACK AND PARTIALLY DISLODGE THE SWIVEL WYE FROM THE ELBOW. THIS MAY HAVE OCCURRED DURING TRANSPORT, WHILE OPENING THE PACKAGING, OR APPLYING EXCESSIVE FORCE WHILE CONNECTING OR DISCONNECTING THE BREATHING CIRCUIT CONNECTORS. ALL BREATHING CIRCUITS ARE PRESSURE TESTED PRIOR TO BEING RELEASED FOR DISTRIBUTION AND ANY CIRCUITS THAT FAIL ARE REJECTED. ANY LEAKS IN THE BREATHING CIRCUIT INCLUDING THE SWIVEL WYE WOULD HAVE BEEN IDENTIFIED DURING THIS TESTING. THIS SUGGESTS THAT THE RETURNED SWIVEL WYE WAS WITHIN SPECIFICATION WHEN RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT267 INFANT EVAQUA2 BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE"
A HOSPITAL IN (B)(6) REPORTED TO FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT SWIVEL WYE-PIECE WAS CRACKED ON TWO RT267 INFANT EVAQUA2 BREATHING CIRCUITS. THESE WERE FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96392 | INFANT EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT267 | 121005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |