SOLETRA
Report
- Report Number
- 3004209178-2014-02863
- Event Type
- Malfunction
- Date Received
- February 12, 2014
- Report Date
- January 21, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- MHY
- PMA / PMN Number
- P960009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
CONCOMITANT: PRODUCT ID 37602, SERIAL# (B)(4), IMPLANTED: 2012-(B)(6), PRODUCT TYPE IMPLANTABLE NEUROSTIMULATOR. PRODUCT ID 37642, SERIAL# (B)(4), PRODUCT TYPE PROGRAMMER, PATIENT. PRODUCT ID 3387S-40, LOT# VA035YD, IMPLANTED: 2012-(B)(6), PRODUCT TYPE LEAD. PRODUCT ID 748351, SERIAL# (B)(4), IMPLANTED: 2012-(B)(6), PRODUCT TYPE EXTENSION, PRODUCT ID 7482A51, SERIAL# (B)(4), IMPLANTED: 2011-(B)(6), PRODUCT TYPE EXTENSION, PRODUCT ID 3387S-40, LOT# V680690, IMPLANTED: 2011-(B)(6), PRODUCT TYPE LEAD.
IT WAS REPORTED THAT THERE WERE LOW OUT OF RANGE IMPEDANCE VALUES. IT WAS NOTED THAT 1 AND 2 MEASURED 29 OHMS. IMPEDANCES WERE C0=964, C1=720, C2=720, C3=1158, 01=982, 02=982, 03=1603, 12=29, 13=1090, AND 23=1084. PATIENT WAS PROGRAMMED AT 2-, 0+ AND THERAPY MEASUREMENT WAS 984 OHMS. IT WAS NOTED THAT WHEN PROGRAMMING EITHER ELECTRODE 1 OR 2 THEY WOULD BE RECRUITING THE OTHER RESPECTIVE ¿SHORTED¿ ELECTRODE. IT WAS FURTHER NOTED THAT THE PATIENT HAD FALLEN ABOUT 2 MONTHS PRIOR TO THIS REPORT. PATIENT HAD TREMOR MANAGEMENT ISSUES, IT WAS NOTED THAT THIS WAS ON-GOING AND NOT RELATED TO THE FALL. IT WAS NOTED THAT THEY HAD BEING ¿TRYING TO GET TREMOR UNDER CONTROL FOR QUITE A WHILE.¿ IT WAS UNKNOWN IF SURGICAL INTERVENTION WOULD BE REQUIRED. ADDITIONAL INFORMATION WAS REQUESTED BUT HAD NOT BEEN RECEIVED AS OF THE DATE OF THIS REPORT. REFERENCE MANUFACTURER¿S REPORT NUMBER: 3004209178-2014-02862.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92034 | SOLETRA | STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR | MHY | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | 7426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00053 YR | Male |