Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 37642, SERIAL# (B)(4), PRODUCT TYPE: PROGRAMMER, PATIENT. PRODUCT ID: 3387S-40, LOT# VA035YD, IMPLANTED: (B)(6) 2012, PRODUCT TYPE: LEAD. PRODUCT ID: 748351, SERIAL# (B)(4), IMPLANTED: (B)(6) 2012, PRODUCT TYPE: EXTENSION. PRODUCT ID: 7482A51, SERIAL# (B)(4), IMPLANTED: (B)(6) 2011, PRODUCT TYPE: EXTENSION. PRODUCT ID: 3387S-40, LOT# V680690, IMPLANTED: (B)(6) 2011, PRODUCT TYPE: LEAD. PRODUCT ID: 7426, SERIAL# (B)(4), IMPLANTED: (B)(6) 2011, PRODUCT TYPE: IMPLANTABLE NEUROSTIMULATOR. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE LOW OUT OF RANGE IMPEDANCE VALUES. IT WAS NOTED THAT 1 AND 2 MEASURED 29 OHMS. IMPEDANCES WERE C0=964, C1=720, C2=720, C3=1158, 01=982, 02=982, 03=1603, 12=29, 13=1090, AND 23=1084. PATIENT WAS PROGRAMMED AT 2-, 0+ AND THERAPY MEASUREMENT WAS 984 OHMS. IT WAS NOTED THAT WHEN PROGRAMMING EITHER ELECTRODE 1 OR 2 THEY WOULD BE RECRUITING THE OTHER RESPECTIVE ¿SHORTED¿ ELECTRODE. IT WAS FURTHER NOTED THAT THE PATIENT HAD FALLEN ABOUT 2 MONTHS PRIOR TO THIS REPORT. PATIENT HAD TREMOR MANAGEMENT ISSUES, IT WAS NOTED THAT THIS WAS ON-GOING AND NOT RELATED TO THE FALL. IT WAS NOTED THAT THEY HAD BEING ¿TRYING TO GET TREMOR UNDER CONTROL FOR QUITE A WHILE.¿ IT WAS UNKNOWN IF SURGICAL INTERVENTION WOULD BE REQUIRED. ADDITIONAL INFORMATION WAS REQUESTED BUT HAD NOT BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
92032 ACTIVA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37602

Patients

Seq Age Sex Outcome Treatment
1 00053 YR Male