FDA Adverse Event Injury Summary report: N

CVS HYDROGEN PEROXIDE CLEANING & DISINFECTING LENS CARE SYSTEM

MDR report key: 3620828 · Received February 10, 2014

Report

Report Number
2026940-2014-00003
Event Type
Injury
Date Received
February 10, 2014
Date of Event
January 7, 2014
Report Date
May 5, 2014
Manufacturer
KC PHARMACEUTICALS, INC.
Product Code
LPN
PMA / PMN Number
K010559
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ASSAY RESULTS ARE PENDING.

Additional Manufacturer Narrative · 1

(B)(4). AS WE DID NOT RETRIEVE COMPLAINT SAMPLE FOR BARREL LENSE CASE, THE RETURNED COMPLAINT SAMPLE SOLUTION WAS EVALUATED USING A RETAINED SAMPLE OF BARREL LENS CASE. BASED ON MANUFACTURER PH SPECIFICATIONS 6.18 - 6.60 AND 6.37- 6.85 PRE-NEUTRALIZATION AND POST-NEUTRALIZATION RESPECTIVELY. THE PRE-NEUTRALIZATION PH 6.14 AND POST-NEUTRALIZATION PH 6.31 WERE SLIGHTLY OUT OF SPECIFICATION. ALTHOUGH THE PH WERE NOTED TO BE SLIGHTLY OUT OF SPECIFICATIONS, MANUFACTURER STATES THE PH RESULTS ARE AT A LEVEL THAT IS RECOGNIZED AS AN ACCEPTABLE PH RANGE 6.0-8.0 FOR OPHTHALMIC SOLUTION PER US PHARMACOPEIA (USP36/NF31).

Description of Event or Problem · 1

ONE (1) PATIENT (B)(6) STATED THAT HE EXPERIENCED A BURNING SENSATION IN THE EYES AFTER USING MEDICAL DEVICE, (B)(6) HYDROGEN PEROXIDE CLEANING & DISINFECTING LENS CARE SYSTEM.

Description of Event or Problem · 1

THIS IS A FOLLOW UP REPORT. INITIAL REPORT WAS SENT ON 02/10/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88564 CVS HYDROGEN PEROXIDE CLEANING & DISINFECTING LENS CARE SYSTEM LENS CARE SOLUTION LPN KC PHARMACEUTICALS, INC. 107097

Patients

Seq Age Sex Outcome Treatment
1 Other