FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3615003 · Received February 8, 2014

Report

Report Number
3004209178-2014-02030
Event Type
Injury
Date Received
February 8, 2014
Report Date
November 18, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT MEDICAL PRODUCTS: 559445, IMPLANTABLE PACING LEAD, (B)(6) 2006; 419388, IMPLANTABLE PACING LEAD, (B)(6) 2006. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY - DEFIBRILLATOR (CRT-D) REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) EARL IER THAN EXPECTED BY THE CUSTOMER. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83314 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D314TRG

Patients

Seq Age Sex Outcome Treatment
1 00089 YR Hospitalization| R 694958, IMPLANTABLE TACHY LEAD