FDA Adverse Event
Injury
Summary report: N
PROTECTA XT CRT-D
MDR report key: 3615003
·
Received February 8, 2014
Report
- Report Number
- 3004209178-2014-02030
- Event Type
- Injury
- Date Received
- February 8, 2014
- Report Date
- November 18, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT MEDICAL PRODUCTS: 559445, IMPLANTABLE PACING LEAD, (B)(6) 2006; 419388, IMPLANTABLE PACING LEAD, (B)(6) 2006. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY - DEFIBRILLATOR (CRT-D) REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) EARL IER THAN EXPECTED BY THE CUSTOMER. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 83314 | PROTECTA XT CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D314TRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00089 YR | Hospitalization| R | 694958, IMPLANTABLE TACHY LEAD |