DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2014-00688
- Event Type
- Other
- Date Received
- February 7, 2014
- Date of Event
- January 10, 2014
- Report Date
- January 10, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV
- Reporter Occupation
- NURSE
Narratives
ON (B)(6) 2014, THE ISI FIELD SERVICE ENGINEER EVALUATED THE PSM ARM 2 AND DID NOT FIND ANY ISSUES; HOWEVER, THE PSM ARM 2 WAS REPLACED DUE TO THE CUSTOMER'S REPORTED COMPLAINT. THE DA VINCI SYSTEM WAS CHECKED AND VERIFIED TO BE WITHIN ISI SPECIFICATIONS. ON (B)(6) 2014, THE PSM ARM 2 INVOLVED WITH THE REPORTED COMPLAINT WAS EVALUATED BY ISI FAILURE ANALYSIS. THE REPORTED ISSUE WAS NOT REPLICATED; HOWEVER, INVESTIGATION NOTED THAT THE PSM ARM 2 FAILED THE BRAKE TEST. AS OF (B)(6) 2014, THE SITE HAS NOT HAD ANY RECURRING ISSUES WITH THE PSM ARM 2. THERE WAS NO REPORT OR ALLEGATION OF PATIENT INJURY; HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE DA VINCI SURGERY WAS CONVERTED TO AN OPEN SURGICAL PROCEDURE AFTER ENCOUNTERING RESISTANCE ISSUES WITH PSM ARM 2.
ON (B)(6) 2014, THE SITE REPORTEDLY ENCOUNTERED ISSUES WITH THE PATIENT SIDE MANIPULATOR (PSM) ARM 2 DURING A DA VINCI PULMONARY LOBECTOMY PROCEDURE. THE SURGEON INFORMED THE REPORTER THAT WHEN HE WAS MOVING THE INSTRUMENTS IN OR OUT, THE MOVEMENT WAS NOT A SMOOTH GLIDING MOTION. THE SURGEON WOULD TRY TO PULL TOWARDS THE TISSUE, IT WOULD NOT MOVE, AND THEN IT WOULD HAVE A JABBING MOTION; HOWEVER, THE PATIENT DID NOT SUSTAIN ANY INJURIES. SHE CLAIMED THAT THE SURGICAL STAFF CHECKED THE PSM ARMS TO MAKE SURE THEY WERE NOT POSITIONED IN A WAY THAT WOULD CREATE RESISTANCE BETWEEN THE PSM'S. THE SURGEON DEALT WITH THE PSM RESISTANCE ISSUE WHILE DISSECTING TISSUE; HOWEVER, WHEN THE SURGEON WAS MOVING CLOSER TO VESSELS APPROXIMATELY 20-30 MINUTES INTO THE PROCEDURE, HE DECIDED TO CONVERT TO OPEN DUE TO THE PSM ARM 2 RESISTANCE ISSUE. THERE WERE NO REPORTED SURGICAL COMPLICATIONS OR PATIENT INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 79506 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000, A6.0P8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |