FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3612555 · Received February 7, 2014

Report

Report Number
0001811755-2014-00394
Event Type
Malfunction
Date Received
February 7, 2014
Date of Event
January 9, 2014
Report Date
January 13, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZK
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND WITHIN THE DEVICE, THE MOTOR PINS WERE DAMAGED, A BEARING WAS STUCK ON THE ROTOR, THE DRIVESHAFT WAS STICKING, AND THE ROTOR COUPLER AND SPINDLE HOUSING WERE WORN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY THE CORE IMPACTION DRILL WAS OVERHEATING. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.