FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3612555
·
Received February 7, 2014
Report
- Report Number
- 0001811755-2014-00394
- Event Type
- Malfunction
- Date Received
- February 7, 2014
- Date of Event
- January 9, 2014
- Report Date
- January 13, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZK
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND WITHIN THE DEVICE, THE MOTOR PINS WERE DAMAGED, A BEARING WAS STUCK ON THE ROTOR, THE DRIVESHAFT WAS STICKING, AND THE ROTOR COUPLER AND SPINDLE HOUSING WERE WORN.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY THE CORE IMPACTION DRILL WAS OVERHEATING. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.