FDA Adverse Event Malfunction Summary report: N

MEDRAD VERIS MR VITAL SIGNS MONITOR

MDR report key: 3612438 · Received December 6, 2013

Report

Report Number
2520313-2013-00057
Event Type
Malfunction
Date Received
December 6, 2013
Date of Event
May 1, 2013
Report Date
May 1, 2013
Manufacturer
MEDRAD
Product Code
MWI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BAYER R & I QA PRODUCT ANALYSIS RECEIVED AND EXAMINED THE RETURNED VERIS MAIN PCB (PRINTED CIRCUIT BOARD) AND DETERMINED IT TO BE MALFUNCTIONING. BAYER R & I SERVICE REPLACED THE PCB MAIN BOARD AT THE SITE AND CONFIRMED SYSTEM FUNCTIONALITY. ON (B)(4) 2013, BAYER HEALTHCARE DISTRIBUTED A FIELD SAFETY NOTICE RECALLING MAIN BOARDS WITH PART NUMBER 301641 THAT WERE INSTALLED IN SOME MEDRAD VERIS MR VITAL SIGNS MONITORS. THESE MAIN BOARDS ARE BEING RECALLED DUE TO THE POTENTIAL FOR UNEXPECTED SHUTDOWN OF THE SYSTEM WHILE IN USE.

Description of Event or Problem · 1

A BAYER R & I REP REPORTED THE FOLLOWING: THE MONITOR WOULD TURN OFF AFTER A FEW MINUTES. THERE WAS NO INJURY OR ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
634859 MEDRAD VERIS MR VITAL SIGNS MONITOR MONITOR, PHYSIOLOGICAL PATIENT MWI MEDRAD 3014268

Patients

Seq Age Sex Outcome Treatment
1