FDA Adverse Event
Other
Summary report: N
HEMOCUE GLUCOSE 201 MICROCUVETTE
MDR report key: 3608638
·
Received November 21, 2013
Report
- Report Number
- 3003044483-2013-00026
- Event Type
- Other
- Date Received
- November 21, 2013
- Date of Event
- October 25, 2013
- Report Date
- October 25, 2013
- Manufacturer
- HEMOCUE AB
- Product Code
- LFR
- PMA / PMN Number
- K20935
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
IN AN INTERNAL INSPECTION, A PRODUCTION RELATED PROBLEM WAS IDENTIFIED DAMAGING THE OUTER PACKAGING MATERIAL OF MICROCUVETTES (POUCH). THE PROBABILITY OF DAMAGED POUCHES REACHING CUSTOMERS IS HIGHLY UNLIKELY BASED ON VISUAL INSPECTION AND REMOVAL OF DAMAGED POUCHES. HOWEVER, DAMAGED POUCHES HAVE REDUCED RESISTANCE TO MOISTURE WHICH AFFECTS THE PERFORMANCE OF THE MICROCUVETTES.
Description of Event or Problem · 1
A PRODUCTION RELATED PROBLEM WAS IDENTIFIED POSSIBLY DAMAGING THE OUTER PACKAGING MATERIAL OF MICROCUVETTES (POUCH). DAMAGED POUCHES CAN ALLOW ENTERING OF MOISTURE AND NEGATIVELY AFFECTING THE PERFORMANCE OF THE MICROCUVETTE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 606792 | HEMOCUE GLUCOSE 201 MICROCUVETTE | LFR GLUCOSE TEST SYSTEM | LFR | HEMOCUE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |