FDA Adverse Event Other Summary report: N

DCA VANTAGE ANALYZER

MDR report key: 3608252 · Received January 23, 2014

Report

Report Number
1217157-2013-00263
Event Type
Other
Date Received
January 23, 2014
Date of Event
December 27, 2013
Report Date
December 27, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Code
CGX
PMA / PMN Number
K071466
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER IS BEING INSTRUCTED TO USE CORRECT SAMPLE TECHNIQUE AND ADVISED TO RUN PRECIOUS STUDY. CUSTOMER RAN PRECISION STUDY AND RESULTS WERE IN RANGE. THE EVENT HAS OCCURRED DUE TO AN OPERATOR ERROR. INSTRUMENT IS PERFORMING AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED DISCORDANT HEMOGLOBIN A1C (HBA1C) RESULTS ON THE INSTRUMENT. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54683 DCA VANTAGE ANALYZER DCA CGX SIEMENS HEALTHCARE DIAGNOSTICS INC

Patients

Seq Age Sex Outcome Treatment
1