FDA Adverse Event Malfunction Summary report: N

ARCHITECT IPTH

MDR report key: 3604666 · Received February 3, 2014

Report

Report Number
3002809144-2014-00015
Event Type
Malfunction
Date Received
February 3, 2014
Report Date
January 13, 2014
Manufacturer
ABBOTT GERMANY
Product Code
CEW
PMA / PMN Number
K063232
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST 8K25-25, THAT HAS A SIMILAR US PRODUCT DISTRIBUTED IN THE US, LIST 8K25-27. (B)(4). A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. AN EVALUATION IS IN PROCESS.

Description of Event or Problem · 1

THE ACCOUNT GENERATED FALSELY ELEVATED ARCHITECT IPTH ON PATIENT SAMPLES THAT TESTED 50% LOWER BY ANOTHER METHOD. EXAMPLES WERE PROVIDED FOR 11 PATIENTS. NO UNIT OF MEASURE WAS PROVIDED. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. DATA PROVIDED: ROCHE METHOD = 31.4, ARCHITECT IPTH = 59.2, ID (B)(4), ROCHE METHOD = 37.1, ARCHITECT IPTH = 64.2, ID (B)(4), ROCHE METHOD = 54.3, ARCHITECT IPTH = 89.0, ID (B)(4), ROCHE METHOD = 52.4, ARCHITECT IPTH= 103.1, ID (B)(4), ROCHE METHOD = 30.3, ARCHITECT IPTH= 55.3, ID (B)(4), ROCHE METHOD = 29.4, ARCHITECT IPTH= 45.9, ID (B)(4), ROCHE METHOD = 33.1, ARCHITECT IPTH = 64.8, ID (B)(4), ROCHE METHOD = 81.8, ARCHITECT IPTH = 151.6, ID (B)(4), ROCHE METHOD = 75.7, ARCHITECT IPTH = 139, ID (B)(4), ROCHE METHOD = 35.3, ARCHITECT IPTH = 67.2, ID (B)(4), ROCHE METHOD = 17.5, ARCHITECT IPTH = 28.0, ID (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
70313 ARCHITECT IPTH CEW ABBOTT GERMANY 01113C000

Patients

Seq Age Sex Outcome Treatment
1 SERIAL (B)(4)| ARCHITECT I1000SR, LIST 01L86-01