ACCOLADE HFX SIZE #5
Report
- Report Number
- 0002249697-2014-00249
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 9, 2014
- Report Date
- January 9, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LZO
- PMA / PMN Number
- K051741
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVENT REGARDING OFF LABEL USE INVOLVING AN ACCOLADE HFX WAS REPORTED. THE DEVICE WAS RETURNED. THE ACCOLADE HFX STEM WAS RETURNED FOR ANALYSIS. THE BULLET TIP OF THE SPECIALTY OFFSET ACCOLADE STEM INSERTER WAS RETURNED FIRMLY STUCK IN THE STEM. IT APPEARED THE TIP OF THE INSERTER WAS ONLY INSERTED INTO THE STEM ABOUT HALF WAY. THERE ARE A FEW MARKS ON THE DISTAL PORTION OF THE STEM. CONCLUSION: IT WAS REPORTED THAT WHILE IMPACTING AN ACCOLADE HFX STEM WITH THE SPECIALTY ACCOLADE OFFSET STEM INSERTER THE INSTRUMENT BECAME LODGED AND STUCK IN THE STEM. A REVIEW OF THE SURGICAL PROTOCOL FOR THE ACCOLADE SYSTEM AND A REVIEW OF THE SURGEON APPROVAL LETTER FOR THE SPECIALTY ACCOLADE OFFSET STEM INSERTER INDICATED THIS INSTRUMENT IS NOT TO BE USED WITH THE ACCOLADE HFX STEM; THEREFORE, THIS APPLICATION OF THE DEVICE IS OFF LABEL USE. NO FURTHER INVESTIGATION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE TO INDICATE FURTHER EVALUATION IS WARRANTED, THIS RECORD WILL BE REOPENED.
WHEN COMPLETED, THE INVESTIGATION RESULTS WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
WHILE IMPACTING AN ACCOLADE HFX STEM, A CUSTOM BROACH HANDLE WAS USED AND BECAME LODGED AND STUCK IN THE THE STEM DURING IMPACTION. WE OPENED ANOTHER STEM AND USED A THREAD IN IMPACTOR WITHOUT ANY FURTHER ISSUES.
WHILE IMPACTING AN ACCOLADE HFX STEM, A CUSTOM BROACH HANDLE WAS USED AND BECAME LODGED AND STUCK IN THE THE STEM DURING IMPACTION. WE OPENED ANOTHER STEM AND USED A THREAD IN IMPACTOR WITHOUT ANY FURTHER ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 70098 | ACCOLADE HFX SIZE #5 | IMPLANT | LZO | STRYKER ORTHOPAEDICS-MAHWAH | MLEWDK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |