SPINEOLOGY OPTIMESH TRAY

MDR report key: 3603245 · Received January 29, 2014

Report

Report Number
MW5034122
Event Type
Injury
Date Received
January 29, 2014
Date of Event
January 9, 2014
Report Date
January 24, 2014
Manufacturer
SPINEOLOGY, INC.
Product Code
GZX
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MD, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING SET-UP IN OPERATING ROOM FOR A MINIMALLY INVASIVE SPINE PROCEDURE, DEBRIS WAS FOUND ON OPERATING ROOM TABLE. DEBRIS CAME OUT OF AN INSTRUMENT FROM A VENDOR LOANED SET. ROOM WAS CLEANED. NEW SUPPLIES AND INSTRUMENTATION WAS OBTAINED. THE THREE PIECES OF EQUIPMENT WERE REMOVED FROM SERVICE AND RETURNED TO VENDOR SINCE IT WAS LOANED EQUIPMENT. ANOTHER TRAY WAS SET UP FOR THE CASE THAT DID NOT HAVE ANY VISIBLE DEBRIS. AS PART OF INVESTIGATION, A STERILE TRAY WAS EXAMINED AND DEBRIS WAS FOUND AFTER MANIPULATING THE EQUIPMENT. THIS MAY BE A PROBLEM WITH THE DESIGN OF THE EQUIPMENT THAT WILL NOT ALLOW YOU TO TAKE IT APART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
62478 SPINEOLOGY OPTIMESH TRAY INSERTER ARTICULATING CURETTE GZX SPINEOLOGY, INC. 3110100
62479 SPINEOLOGY OPTIMESH TRAY ELOCKING MESH HOLDER GZX SPINEOLOGY, INC. 311088
62480 SPINEOLOGY OPTIMESH TRAY 6.5MM SHAPER GZX SPINEOLOGY, INC. 3110042

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention