FDA Adverse Event Summary report: N

60" CENTURY STERILIZER

MDR report key: 3601633 · Received January 31, 2014

Report

Report Number
3005899764-2014-00007
Date Received
January 31, 2014
Date of Event
January 10, 2014
Report Date
January 31, 2014
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE UNIT AND FOUND THAT THE INTERNAL BELLOWS OF THE PRESSURE REGULATORY VALVE (PRV) WERE WORN WHICH CAUSED IT TO RUPTURE AND ALLOW THE STEAM TO ESCAPE. THE TECHNICIAN REBUILT THE PRV, ENERGIZED THE STEAM VALVE, BROUGHT THE UNIT UP TO PRESSURE, RAN A DART WARM-UP TEST AND CONFIRMED THE UNIT TO BE OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT STEAM WAS LEAKING FROM THEIR STERILIZER SUBSEQUENTLY SETTING OFF THE FIRE ALARM AND DISPATCHING THE FIRE DEPARTMENT. THE HALLWAY IN WHICH THE FIRE ALARM IS LOCATED WAS EVACUATED. NO INJURIES OR PROCEDURAL DELAYS/CANCELLATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67039 60" CENTURY STERILIZER STERILIZER FLE STERIS MEXICO, S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment