FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO

MDR report key: 3599415 · Received November 21, 2013

Report

Report Number
2242352-2013-01601
Event Type
Malfunction
Date Received
November 21, 2013
Date of Event
August 20, 2013
Report Date
October 31, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
2242352-10/28/13-002R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVAL. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE BOOTS DISPLAYED EVIDENCE OF HEAT RELATED DAMAGE CONSISTENT WITH ACTIVATION OF THE DEVICE WHILE THE JAWS ARE IN DIRECT CONTACT. A PRE-CAUTERY TEST WAS PERFORMED ON THE DEVICE WITH A REFERENCE POWER SUPPLY AND EXTENSION CABLE. THE DEVICE DID NOT PASS THE PRE-CAUTERY TEST, IT STAYED 'ON'. THE HANDLE ON THE DEVICE WAS OPEN TO VERIFIED CONNECTIONS; UNDER MAGNIFICATION, SOLDERING FLUX WAS OBSERVED ON THE SWITCH BODY, LEVEL AND ACTUATOR OF THE MICRO-SWITCH. THE FLUX CAUSED THE MICRO-SWITCH ACTUATOR TO BE STUCK IN THE 'ON' POSITION. BASED UPON THE EVAL RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED. THE FOLLOWING CORRECTIVE ACTION HAS BEEN TAKEN TO ADDRESS THE SOLDER FLUX ISSUE: MAQUET HAS REVISED THE WORK INSTRUCTION FOR THE SOLDERING PROCESS (B)(4)/HANDLE ASSEMBLY (B)(4) TO ADD ENHANCE VISUAL INSPECTION TO THE SOLDERING POINTS AND THE SWITCH CONTAMINATION WITH FLUX. MAQUET CARDIOVASCULAR LLC HAS INITIATED RECALL (B)(4) TO ADDRESS THIS FAILURE MODE. THE FDA (B)(4) DISTRICT RECALL COORDINATOR HAS BEEN ADVISED. A NOTIFICATION LETTER HAS BEEN SENT TO THIS CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO REMAINED ACTIVATED AFTER THE ACTIVATION TOGGLE WAS RELEASED. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607696 VASOVIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25078104

Patients

Seq Age Sex Outcome Treatment
1 NI