ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00085
- Event Type
- Malfunction
- Date Received
- January 29, 2014
- Date of Event
- January 8, 2014
- Report Date
- January 10, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR INSPECTION. IT WAS PRESSURE TESTED, VISUALLY INSPECTED, AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST RESULT WAS OUT OF SPECIFICATION DUE TO EXCESSIVE LEAKAGE COMING FROM THE PATIENT END CONNECTOR. THIS WAS CONFIRMED WHEN THE SUBJECT BREATHING CIRCUIT WAS IMMERSED IN THE WATER BATH. VISUAL INSPECTION REVEALED THAT THERE WAS NOT ENOUGH GLUE PRESENT IN THE EVAQUA EXPIRATORY LIMB CONNECTOR, CAUSING THE REPORTED LEAK. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130924. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED BEFORE RELEASING FOR DISTRIBUTION, AND THOSE THAT FAIL ARE REJECTED. THE OBSERVED LEAK WAS CAUSED BY INSUFFICIENT GLUE BEING INJECTED INTO THE EVAQUA EXPIRATORY LIMB CONNECTOR SUCH THAT IT DID NOT FORM A PERMANENT SEAL DURING THE ASSEMBLY PROCESS. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS". OUR MONITORING AND TRENDING OF COMPLAINTS INVOLVING INCORRECTLY ASSEMBLED ADULT EVAQUA BREATHING CIRCUITS DUE TO INSUFFICIENT GLUE APPLIED HAS A RATE OF OCCURRENCE OF (B)(4) DEVICES PER (B)(4) SOLD WORLDWIDE IN THE LAST YEAR TO THE END OF DECEMBER 2013.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED BEFORE USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 63216 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LIMITED | RT340 | 130924 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |