FDA Adverse Event Malfunction Summary report: N

ZIMMER SKIN GRAFT MESHER

MDR report key: 3597708 · Received November 22, 2013

Report

Report Number
1526350-2013-00702
Event Type
Malfunction
Date Received
November 22, 2013
Date of Event
October 1, 2013
Report Date
October 13, 2013
Manufacturer
ZIMMER SURGICAL
Product Code
FZW
Removal / Correction Number
RES 61798
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

BEGINNING MAY 31, 2012, U.S. AND CANADA CUSTOMERS WERE SENT AN URGENT PT SAFETY ADVISORY INFORMING THEM THE NEED FOR PROPER CARE AND PREVENTIVE MAINTENANCE OF THE ZIMMER SKIN GRAFT MESHER. CUSTOMER WERE INFORMED THAT IMPROPERLY MAINTAINED INSTRUMENTS MAY CAUSE DONOR SITE INJURIES OR RESULT IN DAMAGE TO THE GRAFT. CUSTOMERS WERE REQUESTED TO CONTACT ZIMMER TO SCHEDULE MAINTENANCE FOR THE DEVICE IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND EVAL. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 12/15/1992 AND WAS LAST REPAIRED ON (B)(4) 2012 FOR A NON-RELATED ISSUE. EVAL OF THE DEVICE OBSERVED A COMPLETELY SHEARED AND BENT RIGHT TIME ON THE COMB, AS WELL AS A BENT REAR LIP ON THE COMB. WEAR TO BOTH THE RIGHT END OF THE ROLLER AND ROLLER GEAR WAS ALSO OBSERVED PRIOR TO REPAIR, A TEST MESH AND CALIBRATION CHECK COULD NOT BE PERFORMED DUE TO THE DAMAGED COMB. CUSTOMER DID NOT RETURN A RATCHET OR CUTTERS FOR EVAL. THE CAUSE IS LIKELY THE USER NOT MAINTAINING THE DEVICE PER PROPER HANDLING AND LACK OF PREVENTATIVE MAINTENANCE. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE COMB OF THE ZIMMER SKIN GRAFT MESHER WAS BENT. NO ADDITIONAL CLINICAL INFO WAS RECEIVED PRIOR TO THIS REPORT. A FOLLOW UP MEDWATCH WILL BE SUBMITTED IF ADDITIONAL CLINICAL INFO IS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
611099 ZIMMER SKIN GRAFT MESHER ZIMMER SKIN GRAFT MESHER FZW ZIMMER SURGICAL NA NA

Patients

Seq Age Sex Outcome Treatment
1